Search papers, labs, and topics across Lattice.
This paper introduces a novel perception and control framework for autonomous endovascular navigation using 2D image feedback, converting vessel masks into a labeled centerline graph and projecting device-tip estimates onto a planned path to derive navigation metrics. A rule-based controller then leverages these metrics for complex maneuvers like branch selection and recovery. The method achieved high success rates (96.9%) in closed-loop runs on a physical robotic platform, demonstrating its potential for image-guided robotic interventions.
Autonomous endovascular navigation achieves 96.9% success on a physical robotic platform using only 2D X-ray fluoroscopy feedback.
In endovascular interventions, physicians rely on planar DSA and X-ray fluoroscopy to interpret vessel anatomy and device motion, yet converting these images into feedback for autonomous navigation remains difficult. A perception and control method is formulated for two-dimensional image feedback. Vessel masks are converted into a labeled centerline graph, and device-tip estimates are projected onto a planned path to obtain progress, deviation, and direction. A rule-based controller uses these quantities for delivery, branch selection, probing, rollback, and recovery. The method was evaluated on a physical dual-device robotic platform in 65 closed-loop runs using simulator-rendered planar images across three paths. All runs reached at least 0.95 maximum observed guidewire progress; two runs (3.1%) met the prespecified failure rule of three or more autonomous stop events. The results establish route-level performance on the tested platform; clinical image sequences and broader anatomy remain to be evaluated.