Search papers, labs, and topics across Lattice.
The paper introduces an open-source hardware-software architecture designed to stabilize real-time multi-spectral photoacoustic imaging (RT-mPAI) systems using OPOTEK Phocus lasers and Verasonics Vantage platforms. They achieve this by implementing a microcontroller for precise laser trigger management and a decoupled client-server data streaming framework, mitigating timing deviations caused by non-real-time operating systems. The result is a democratized, stable RT-mPAI system with reduced technical and cost barriers, fostering broader research accessibility.
Achieve stable real-time multi-spectral photoacoustic imaging with commodity hardware by sidestepping OS timing issues using a clever microcontroller-based synchronization scheme.
Real-time multi-spectral photoacoustic imaging (RT-mPAI) often suffers from synchronization instabilities when interfacing fast-tuning lasers with data acquisition platforms executing on non-real-time operating systems. To overcome this, we establish an open-source hardware-software architecture tailored for the widely adopted combination of the OPOTEK Phocus lasers and Verasonics Vantage systems. By employing an independent micro-controller for deterministic laser trigger counting alongside a decoupled client-server data streaming framework, the proposed system circumvents OS-induced timing deviations and local storage bottlenecks. By open-sourcing this pipeline and cultivating a collaborative environment to share both code and ideas, we aim to lower the technical and cost barriers for RT-mPAI, thereby democratizing access to stable RT-mPAI research and, more ambitiously, fostering a vibrant open-source community.