This paper addresses the challenge of safe control for steerable catheters by modeling the interaction dynamics between the catheter tip and tissue, focusing on maintaining compliance and adhering to strict contact-force limits. A predictive optimizer is employed to regulate the interaction state while respecting constraints on contact force, tendon force, and curvature, resulting in significant improvements in tracking accuracy and safety. The findings reveal that effective motion regulation and contact-force safety are interconnected objectives, with the proposed method achieving a 90% reduction in approach error while maintaining contact force within clinically relevant bounds.
Achieving a 90% reduction in tracking error while ensuring catheter contact force remains safely below critical thresholds could revolutionize steerable catheter applications in medical procedures.
Safe steerable catheter control is fundamentally a problem of interaction dynamics: the tip must follow a planned motion, remain compliant against moving tissue, reject friction and hysteresis, and respect a clinically meaningful never-exceed contact-force bound. We formulate catheter--tissue interaction dynamics in the scalar tip-normal coordinate of a single-segment single-tendon catheter. A partial-physics feedforward cancels only the reliable nominal bending dynamics, exposing a configuration-invariant linear interaction-dynamics model whose input gain varies through the scalar catheter inertia. A predictive optimizer then regulates this interaction state subject to hard contact-force, tendon-force, and curvature constraints. An augmented Kalman filter compresses contact, friction, and modeling error into one sensor-free disturbance state, giving nominal offset-free regulation in free space while leaving force safety to the explicit constraint. The unconstrained and disturbance-free limit recovers classical catheter impedance as a special realization of the same interaction dynamics, rather than as the main design object. In a MuJoCo distributed-compliance simulation of an eight-link tendon-driven catheter, disturbance augmentation cuts free-space approach error by 90\%, and only the force-constrained predictive interaction-dynamics controller reconciles tracking with the 0.5\,N bound: the unconstrained controller drives contact force to 0.60\,N against a penetrating target, while the constrained one holds 0.47\,N at identical tracking. These results show that offset-free motion regulation and contact-force safety are coupled interaction-dynamics objectives, and that the explicit predictive constraint resolves their tension under stiff tissue contact. The bound also holds under $0.5$\,mm, $1.2$\,Hz cardiac motion. Hardware validation is future work.